Site Preparations
3-8 Ultrasound SystemCommon Service Information
Direction 5444964-100 English
Rev. 5
Electrical requirements for Console Ultrasound Systems (continued)
The following power line parameters should be monitored for
one week before installation. We recommend that you use an
analyzer Dranetz Model 606-3 or Dranetz Model 626:
Table 3-4: Electrical Specifications for LOGIQ C Series and LOGIQ C3/C5 Premium
Parameter Area Limits
Voltage Range 100-120V~ 500VA
220-240V~ 500VA
Power All applications MAX. 750 VA
Line Frequency All applications 50/60Hz (±2Hz)
Power Transients All applications Less than 25% of nominal peak
voltage for less than 1 millisecond
for any type of transient, including
line frequency, synchronous,
asynchronous, or aperiodic
transients.
Decaying Oscillation All applications Less than 15% of peak voltage
for less than 1 millisecond.

Table of Contents

Other manuals for GE LOGIQ S8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ S8 and is the answer not in the manual?

GE LOGIQ S8 Specifications

General IconGeneral
TypeUltrasound System
Display TypeLCD
3D/4D ImagingYes
ConnectivityDICOM, USB, Ethernet
Display Size19 inches
TransducersLinear, Convex, Phased Array
Doppler ModesColor Doppler, Power Doppler, Spectral Doppler
Operating ModesB-Mode, M-Mode
ApplicationsAbdominal, Cardiac, Obstetrics, Gynecology, Vascular, Musculoskeletal
PortabilityCart-based
Transducer Ports4

Related product manuals