Site Preparations
3-14 Ultrasound SystemCommon Service Information
Direction 5444964-100 English
Rev. 5
Electrical Requirements for Isolation Cart
Table 3-11: Isolation Cart for LOGIQBook XP Series/LOGIQ e/LOGIQ e Vet/LOGIQ i/
Vivid e
Input Output
100-120V~, 50/60Hz, 500VA 100-120V~, 50/60Hz, 250VA
220-240V~, 50/60Hz, 500VA 220--240V~, 50/60Hz, 250VA
Table 3-12: E-Isolation Cart for LOGIQ e
Input Output
100-120V~, 60Hz, 400VA 100-120V~, 60Hz, 350VA
220-240V~, 50Hz, 400VA 220--240V~, 50Hz, 350VA
Table 3-13: Advanced Isolation Cart
Input Output
100-120V~, 60Hz, 400VA 100-120V~, 60Hz, 350VA
220-240V~, 50Hz, 400VA 220--240V~, 50Hz, 350VA