General Ultrasound system requirements
Ultrasound SystemCommon Service Information 3-7
Direction 5444964-100 English Rev. 5
Table 3-3: Electrical requirements - LOGIQ Consoles
Product Voltage Tolerance
Current or
Power
Consumption
Frequency
LOGIQ E10 100-240 VAC
±10%
0.9kVA
50/60Hz (±2Hz)
LOGIQ E9 100-240 VAC 1100 W (valid
through R5)
1.0kVA (effective
from R6)
LOGIQ P5
LOGIQ A5
100-120 VAC
200-240 VAC
750 VA
LOGIQ E8
LOGIQ S8
LOGIQ S7
100-120 VAC
200-240 VAC
900 VA
LOGIQ P3 100-120 VAC
200-240 VAC
425 VA
LOGIQ P6 100-120 VAC
200-240 VAC
950 VA
LOGIQ P9/P7 100-120 VAC
200-240 VAC
500 VA
LOGIQ 9 100-120 VAC 9.5 A
200-240 VAC 4.75 A
LOGIQ 7
100 VAC
MAX. 1200 VA
115 VAC
220 VAC
LOGIQ 5
100 VAC
MAX. 1250 VA
115 VAC
230 VAC
LOGIQ 3
100 VAC
MAX. 860VA
115 VAC
230 VAC
LOGIQ F Series 100-240 VAC
±10%
MAX. 400 VA
50/60Hz (±2Hz)
LOGIQ C9 Series 220-240 VAC MAX. 950 VA
LOGIQ V3/LOGIQ
V5/LOGIQ V5
Expert
100-240 VAC MAX. 300 VA

Table of Contents

Other manuals for GE LOGIQ V2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ V2 and is the answer not in the manual?

GE LOGIQ V2 Specifications

General IconGeneral
BrandGE
ModelLOGIQ V2
CategoryMedical Equipment
LanguageEnglish

Related product manuals