EasyManuals Logo
Home>GE>Medical Equipment>LOGIQ V2

GE LOGIQ V2 Technical Publication

GE LOGIQ V2
127 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #60 background imageLoading...
Page #60 background image
Site Preparations
3-8 Ultrasound SystemCommon Service Information
Direction 5444964-100 English
Rev. 5
Electrical requirements for Console Ultrasound Systems (continued)
The following power line parameters should be monitored for
one week before installation. We recommend that you use an
analyzer Dranetz Model 606-3 or Dranetz Model 626:
Table 3-4: Electrical Specifications for LOGIQ C Series and LOGIQ C3/C5 Premium
Parameter Area Limits
Voltage Range 100-120V~ 500VA
220-240V~ 500VA
Power All applications MAX. 750 VA
Line Frequency All applications 50/60Hz (±2Hz)
Power Transients All applications Less than 25% of nominal peak
voltage for less than 1 millisecond
for any type of transient, including
line frequency, synchronous,
asynchronous, or aperiodic
transients.
Decaying Oscillation All applications Less than 15% of peak voltage
for less than 1 millisecond.

Table of Contents

Other manuals for GE LOGIQ V2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ V2 and is the answer not in the manual?

GE LOGIQ V2 Specifications

General IconGeneral
BrandGE
ModelLOGIQ V2
CategoryMedical Equipment
LanguageEnglish

Related product manuals