EasyManua.ls Logo

GE MAC 1200

GE MAC 1200
151 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Revision D MAC 1200 resting ECG analysis system 2-9
2012250-095
Functional Description: PCB Control CS_CI
ECG Recording and Front-End Processing
The patient input is classified as being cardiac floating and is defibrillation-proof.
The patient leads are connected with a 15-pin connector with a special form for the
specified and released patient cables for the MAC 1200 resting ECG analysis
system.
The main element of the ECG recording and front-end processing is a set of chips
comprising 3 ASICs. Pace detection is realized by a separate circuit.
Every disconnected electrode is detected by a special AC measurement, which
allows higher impedance between electrode and patient.
N common-mode compensation ensures suppression of interference, at the same
time serving to improve the in-phase suppression of the input electrodes.
To protect patients, the ECG recording and front-end processing are assembled as
floating components. Digital signals from and to the controller core are transmitted
via opto-electronic couplers. The floating supply ±5 V is generated by an isolated
flyback converter from the +5 V logic power supply.

Table of Contents

Other manuals for GE MAC 1200

Related product manuals