14693e – 31 Mar 03 1 - 12 General Information
1.10 Specifications:
1.10.1 Input Water Requirements
Minimum Maximum
Input flow Rate 2.0 GPM N/A
Temperature (water) 1.7°C (35°F) 32°C (90°F)
NOTE:
Operating at lower temperatures may make it impossible for the Millenium™ to
produce adequate permeate for the dialysis machine.
pH (operating) 4.0 11.0
TF Membranes
pH w/chloramines present (operating) 4.0 8.5
TF Membranes
pH (cleaning) 2.0 11.0
TF Membranes
Pressure, dynamic 30 PSI 100 PSI
1.10.2 Flow Rates
Product Flow based on the following conditions:
New Membranes
Operating Pressure: 190 PSI
Product Pressure: 1 PSI
Feed Water Temperature: 77°F (25°C)
GPM LPM GPD LPD
.34 1.31 500 1892.5
.52 1.97 750 2839.1
Membrane flux may vary ±15%
Waste Flow: .84 GPM (3.2 LPM) (
±10%)
1.10.3 Electrical Requirements
Motor 1/3 HP
Voltage 115 VAC / 230 VAC
Hertz 60 / 50
Amps 8 / 5
Phase single