There are five kinds of water cooling system configurations.
Type A, B and C have their individual plumbing assembly.
In type B and type E, Facility water is provided by the customer. The water should meet
the specification in the pre installation manual. In not, it is possible to damage our
equipment.
In Type D, the BRM chiller and Air-cooled cryogen compressor are only available to 200V
AC power input. So the facility power can only be 200V AC.
Type E is only for EMEA. A 3 Phase AC stepdowntransformer should be installed which
supplies the 200V to BRM chiller.