D The generator can supply a maximum of 100 W to the tube.
This example represents one 30 kV/200mAs exposure each minute.
2–1–6 Line Conditioner Resonance Frequency
D 20 kHz approx.
2–1–7 Generator Power (excluding tube)
D 5 kW maximum.
2–1–8 Maximum Tolerance of Displayed Constants (with and without AEC or AOP)
D kV: "2% + M1+M1’
D mAs: "5% "1 mAs + M4 + M4’
+$ , " !   " ! 
+ I " ;D "M 9,H ,  !  # "#
+) I " 5 "M 9,H ,
+F I 5, " ! # 
+)F I 5, " ! # 
2–1–9 Measurement Conditions
D kVp: Connect a voltage divider such as the Machlett Dynalizer in
series with the x-ray tube and measure the voltage with an
oscilloscope,
D mAs: Use a waveform analyzer such as the Tektronic 7854, connected
in series with the x-ray tube. The oscilloscope receives a signal
proportional to the high voltage applied to the tube. The mAs
values are given by:
mAs =
ŕ
.
.
i dt
Where T1 is the time at which the high voltage reaches 75% of
its maximum value, and
T2 is the time at which the high voltage returns to 75% of the
maximum value.
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

Other manuals for GE SENOGRAPHE DMR+

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE SENOGRAPHE DMR+ and is the answer not in the manual?

GE SENOGRAPHE DMR+ Specifications

General IconGeneral
BrandGE
ModelSENOGRAPHE DMR+
CategoryMedical Equipment
LanguageEnglish

Related product manuals