Setup
16 TONOPORT V 2001589-085 Revision G
Time and Date
Usually the BP monitors are set to the correct time and
date before delivery. Therefore, the time only needs to be
corrected to change between Standard Time and Daylight
Saving Time.
Setting Date and Time
Briefly switch TONOPORT V off and on again and
wait for the time to be displayed.
Push twice: the display indicates "H 2".
Push : the display indicates the year, e.g.
"2009".
If the indicated year is correct, confirm it with
or correct it with and confirm with .
The display indicates the month, e.g. "03".
If the indicated month is correct, confirm it with
or correct it with and confirm with .
In the same manner, correct day, hour and minute.
In the end, the time of day will be displayed again.
Selecting the Pressure Unit
Briefly switch TONOPORT V off and on again and
wait for the time to be displayed.
Push eight times: the display indicates "H 8".
Push : the display indicates "mmHg" or "kPa".
Confirm the pressure unit with or select the oth-
er unit with , then confirm with .
Measurement Protocols
There is a choice of three different measurement
protocols:
Max. inflation pressure: day phase 250 mmHg
night phase 220 mmHg
Select a Measurement Protocol
Briefly switch TONOPORT V off and on again and
wait for the time to be displayed.
Push three times: the display indicates "H 3".
Push : the display indicates "LLLL" (Selecting
a protocol automatically clears the memory. If you do
not wish to clear the memory, turn the BP monitor
off.)
Push : the display indicates "P1" (protocol 1).
Using , select protocol 2 or 3
OR
Confirm the displayed protocol with .
Enable or Disable the Audio Signal
Briefly switch TONOPORT V off and on again and
wait for the time to be displayed.
Push seven times: the display indicates "H 7".
Push : the display indicates "0000" when the au-
dio signal is turned off, and "1111" when it is turned
on.
Either confirm the setting with or press to
select the alternate setting and confirm with .
START
STOP
INFO
START
STOP
INFO
START
STOP
START
STOP
INFO
START
STOP
START
STOP
INFO
START
STOP
INFO
START
STOP
START
STOP
INFO
START
STOP
START
STOP
INFO
START
STOP
INFO
START
STOP
START
STOP
INFO
START
STOP
INFO
START
STOP
START
STOP
INFO
START
STOP
START
STOP
INFO
START
STOP
INFO
START
STOP
Protocol Day Phase
(7 a.m. to 10 p.m.)
Night Phase
(10 p.m. to 7 a.m.)
P1 every 15 minutes every 30 minutes
P2 every 20 minutes every 40 minutes
P3 every 30 minutes every 60 minutes
START
STOP
INFO
START
STOP
INFO
START
STOP
START
STOP
INFO
START
STOP
START
STOP
INFO
START
STOP
INFO
START
STOP
START
STOP
INFO
START
STOP
INFO
START
STOP
START
STOP
INFO
START
STOP
START
STOP
INFO
START
STOP
INFO
START
STOP

Other manuals for GE TONOPORT V

Related product manuals