Electromagnetic Compatibility (EMC)
2
001589-085 Revision G TONOPORT V 30
11 Appendix - Electromagnetic 
Compatibility (EMC)
Changes or modification to this system not expressly 
approved by GE Healthcare could cause EMC issues with 
this or other equipment This system is designed and 
tested to comply with applicable regulation regarding 
EMC. It needs to be installed and put into service 
according to the EMC information stated as follows.
Warning
Use of portable phones or other radio frequency 
(RF) emitting equipment near the system may cause 
unexpected or adverse operation.
Warning
The equipment or system should not be used 
adjacent to, or stacked with, other equipment. If 
adjacent or stacked use is necessary, the equipment 
or system should be tested to verify normal 
operation in the configuration in which it is being 
used.
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
TONOPORT V is intended for use in the electromagnetic environment specified below. It is the responsibility of the 
customer or user to ensure that TONOPORT V is used in such an environment.
Emissions Test Compliance Electromagnetic Environment - Guidance
RF emissions to EN 55011/
CISPR 11
Group 1 TONOPORT V uses RF energy only for its internal 
function.  Therefore, its RF emissions are very low and 
are not likely to cause any interference in nearby 
electronic equipment.
RF emissions to EN 55011/
CISPR 11
Class B TONOPORT V is suitable for use in all 
establishments, including domestic and those directly 
connected to the public low-voltage power supply 
network that supplies buildings used for domestic 
purposes.
Harmonic emissions to EN 
61000-3-2/IEC 61000-3-2
not applicable
Voltage fluctuations/flicker 
emissions to EN 61000-3-3/IEC 
61000-3-3
not applicable