Cleaning, Maintenance, Disposal
27 TONOPORT V 2001589-085 Revision G
Disposal of the Product Calibration Mode
(e.g. to check the pneumatic system for leaks)
Connect a rubber bulb between pressure tubing and
cuff, using a T-adapter.
Roll up cuff tight.
Switch off device and switch it on again after a few
seconds.
Wait for time to be displayed.
Push four times: the display indicates "H 4".
Push : the display indicates an internal value that
must be between 25 and 100. If the displayed value is
outside this range, TONOPORT V must be returned
for repair.
Push again: the display indicates "0" (the dis-
play now indicates the pressure in mmHg).
Generate a test pressure of 200 mmHg and measure
the pressure decrease after waiting at least 30 sec-
onds. (Pressure decreases between 3 and 5 mmHg are
typical; a pressure decrease > 6 mmHg indicates a
leak and the system needs to be repaired.)
Push to exit the calibration mode.
View Firmware Version
Turn on the device and wait for the time to be dis-
played.
Push five times: the display indicates "H 5".
Push : the firmware version is indicated, e.g.
– "21:00" = firmware version 2.1
Push to end the display.