GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03)
D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
Chapter 4 - Functional Checks 4-35
4-5-4 Load Full Backup (Presets, Configurations & Application Settings)
WARNINGWARNING
!! WARNING:
Always backup any data before an upgrade; see: Section 4-5-3 on page 4-33.
The “Full Backup” loading procedure replaces (overwrites) ALL the existing data on the
local hard drive of the Voluson® 730Expert system!
CAUTION
!! CAUTION:
There are circumstances where it is not possible to load (restore) all the data.
The following rules specify these restrictions:
1.) Generally, only restoring data from an older to a newer software version is possible.
Loading a backup into a system that has a lower software version than the system the backup was
created on is prohibited.
2.) Options can only be restored on the same Voluson® 730Expert system within the same major
software version.
3.) When loading a backup into a system with a software version that has a higher major number
(2.x.x -> 3.x.x -> 4.x.x -> 5.x.x), the following items will not be restored:
A.) User Settings
B.) Options
C.) State of the Service Platform (new model type necessary for VOLC)
4.) The user is only allowed to restore data to a different system if and only if the software version on
this system is the same as in the backup.
5.) The user is only allowed to restore data onto the same system if and only if the software version
on this system is equal or higher than the version in the backup.
6.) The user is not allowed to restore the following items to a different system:
A.) Windows Network Settings
B.) Options
C.) DICOM AE Title
D.) DICOM Station Name
E.) State of the Service Platform