GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03)
D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
Chapter 4 - Functional Checks 4-39
Section 4-6
Software Configuration Checks
Touch the UTILITIES key, then touch SYSTEM SETUP.
The System Setup desktop offers different pages to check:
Touch the UTILITIES
key, then touch MEASURE SETUP.
The Measurement Setup desktop offers different pages to check:
Table 4-10 System Setup Checks
Step Task Expected Result(s)
1
General: Check Date and Time setting Date and Time are correct
2
General: Check that Location (Clinic Name) is correct Location Name is correct
3
General: Check Language settings desired Language is displayed
4
User Setting: Check all the User Settings settings assigned as desired by the customer
5
User Setting: Check Touch Panel Scheme
color scheme display of the Touch Panel assigned as desired
by the customer
6
Peripherals: Check assignment of Printer Keys
Print A and Print B keys are assigned as desired by the
customer
7
Peripherals: Check assignment of Foot Switch Foot Switch left and right are assigned as desired by customer
8
Peripherals: Check Save Destination assignment setting assigned as desired by the customer
9
Peripherals: Check 2D Save key assignment setting assigned as desired by the customer
10
Peripherals: Check Remote Print A, Remote Print B,
Report Printer and DICOM Print Job assignment
settings assigned as desired by the customer
11
Option: Check that all of the options are set up correct all authorized functions are enabled
12
Network:
Check DICOM, Sonoview and Network configuration
setting assigned as desired by the customer
Table 4-11 Measurement Setup Checks
Step Task Expected Result(s)
1
General: Check “2D Circumference and Area Method” setting assigned as desired by the customer
2
General: Check “Show Caliper Information Text” setting assigned as desired by the customer
3
General: Check “Caliper & Measure-Result Font Size” setting assigned as desired by the customer
4
General: Check “Doppler Manual Trace Mode” setting assigned as desired by the customer
5
General: Check “Store Measurements” setting assigned as desired by the customer
6
General: Check “Doppler Trace Display Results” settings assigned as desired by the customer
7
General: Check “Measure Results Display” setting assigned as desired by the customer
8
General: Check “On Freeze Event Start...” setting assigned as desired by the customer
9
OB: Check “Preset Selection” setting assigned as desired by the customer
10
OB: Check “Pregnancy Weeks” setting assigned as desired by the customer
11
OB: Check “Ratio Calculations” settings assigned as desired by the customer
12
OB: Check “GS Measurement Method” setting assigned as desired by the customer