GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03)
D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
4-40 Section 4-7 - Peripheral Checks
Section 4-7
Peripheral Checks
Check that peripherals work as described below:
4-7-1 ECG Check Out
Connect the ECG preamplifier MAN and check:
4-7-2 Power Supply Adjustment
There are no adjustments on the power supplies. The DC Power is self-regulated. If a voltage is outside
the specified range, it means that something is wrong, either with the power supply itself or with a
component connected to that specific power outlet.
13
OB: Check “Show Author´s Name at Measure Menu” setting assigned as desired by the customer
14
Cardiac: Check “2D Circumference and Area Method setting assigned as desired by the customer
15
Cardiac: Check “LV Volume Calculation Method” setting assigned as desired by the customer
Table 4-12 Peripheral Checks
Step Task to do Expected Result(s)
1
Press the FREEZE
key. Stop image acquisition.
2
Press the PRINT A
or PRINT B key on the
Control Panel.
The image displayed on the screen is printed on printer, depending on
the key assignment configuration
3
Press the VCR
key on the Control Panel twice. VCR starts recording (REC - will be displayed on the screen)
4
Press the VCR
key twice again. VCR stops recording
5
Press the VCR
key on the Control Panel once. The VCR Remote Control menu is displayed on the Touch Panel.
6
Press RECORD
on the Touch Panel.
to start recording: A red dot is displayed in the VCR status area on the
Title bar to indicate that recording has begun
7
Press STOP
on the Touch Panel. To Stop recording: The video status icon is changed to (Pause)
8
Press PLAY
on the Touch Panel. To start, Play back an examination
9
Press EXIT
on the Touch Panel to return to the scanning mode
10
Use the assignable keys on the Touch Panel
to perform actions on the recorded session, such as stop, pause,
rewind or fast forward. The video status icon in updated accordingly.
Table 4-13 ECG preamplifier Check
Step Task Expected Result(s)
1
Connect the ECG at the Connector on the rear panel of the
scanner. Press the ECG
key on the control panel to
display the “ECG” menu on the Touch Panel.
It will display a curve along the bottom edge of the image
sector
Table 4-11 Measurement Setup Checks
Step Task Expected Result(s)