7-6 Section 7-4 - Screen Captures and Logs
7-4-2-1 Export System Data
Every time an error message like Figure 7-4 is produced, a dump-file and a text file containing the error
dump and the error message are created in D:\export. Up to 20 dump files are stored there.
NOTE: To deposit this screen as an image file on D:\export, press ALT
+ P.
For more details see: Section 7-4 "Screen Captures and Logs" on page 7-5.
After clicking on OK
the system reboots automatically.
By selecting the EXPORT SYSTEM DATA...
button, these dump-files and text files, the Event Log File,
the full Service Database and information about probes, boards, SW, Options and Service Actions can
be exported to the DVD/CD Drive (or the optional MO Drive).
Figure 7-4 system has encountered a problem
important
dump-file