GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03)
D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
7-22 Section 7-8 - Error Messages
B-DynContrast Out Of Range press ok and save this user-setting once again
BF: can't set BM RxApod restart the system
BF: can´t set C Rx Apod restart the system
BF: can't set D RxApod restart the system
binary file not found restart the system
Binary path not set restart the system
BM_Resample: overrun SampleLengthOnLineMem restart the system
Cannot create instance restart the system
Cannot read a valid Probe-ID (xxx) from…………. disconnect and reconnect the probe
CANNOT_IMPORT_VOLUME_DATA_TO_3D_DLL load volume files from other storage medium
Can't create hardware configuration index file in D:\\SERVICE restart the system
Can't detect PCI 9054 (CRS) restart the system
Can't import session use a new CDR to write data to CD
Can't open file: …… restart the system
can't open MotCtrl RS232 Serial connection restart the system
Can't open: \\P a t h \Load_HW.log restart the system
Can't open: .......... restart the system
Can't open: C:\V730\Distribution\Hardware\patterndat.bin restart the system
can't write XilinxFiles.ini restart the system
CD write error use a new CDR
Cine2D_CtrlBlock::vSetState: value of m_GIP_eCineType
undefined!!
restart the system
Cine2D_CtrlBlock::vSetState: value of m_pGIP_eR_W_Mode
undefined!!
restart the system
Communication thread is dead! restart the system
ConManager - RC_ConstructGOPs: Memory allocation failed restart the system
CPF Hardware not found restart the system
CPS_InterfaceToHW::vSet_VOL_MODE -- incorrect value for
ColorDMA_Active
restart the system
CPS_InterfaceToHW::vSet_VOL_MODE -- incorrect value for
EndOfFrameInt
restart the system
CPS_InterfaceToHW::vSet_VOL_MODE -- incorrect value for
LongLineMemFIFO
restart the system
CPS_InterfaceToHW::vSet_VOL_MODE -- incorrect value for
LongLineMemFIFO
restart the system
Error Messages Actions