EasyManua.ls Logo

GE Voluson 730Expert - Section 7-8 Error Messages

GE Voluson 730Expert
332 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03)
D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-21
Section 7-8
Error Messages
NOTICE
!! NOTICE:
If the problem (error message) still exists after performing the described actions, call technical support.
Error Messages Actions
3D B_RxLines exceeded! restart the system
3D B-O_Frames exceeded! restart the system
3D C_RxLines exceeded! restart the system
3D C-O_Frames exceeded! restart the system
530-Probe connected on……….
Disconnect and reconnect the probe.
If error remains don't use such a probe.
AcousticUnitType not supported by BF_Interface: restart the system
Array bounds exceeded restart the system
Attempt to read Probe-ID from an invalid probe connector. reboot system.
AVI Save function fails
check the SCSI and the Power cable - restart the machine
and try again
B_Enhance Out Of Range press ok and save this user-setting once again
B_Gain Out Of Range press ok and save this user-setting once again
B_Reject Out Of Range press ok and save this user-setting once again
B_TxFocus - not calculated and B_SHOT_PART_ON restart the system
Backup error while writing. (Error during writing of backup data.)
check storage destination for Full Backup
(e.g., DVD/CD not empty, insufficient rights on target
Network drive, write protection on MO or USB-drive)
Backup error while verifying. (Checksum mismatch) repeat backup
BC Ensemble Out Of Range press ok and save this user-setting once again
BC lines_per_sequenz < 1 restart the system
BC lines_per_sequenz < 2 restart the system
BC_Dynamic Out Of Range press ok and save this user-setting once again
BC_Gain Out Of Range press ok and save this user-setting once again
BC_Lines: BC_LineDensity out of limit restart the system
BCMC_Balance Out Of Range press ok and save this user-setting once again
bCP_ConvertParameters failed restart the system
bCP_ProcessIQEnsemblePacket failed restart the system
bCP_ProcessIQMModePacket failed restart the system