GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03)
D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
7-20 Section 7-7 - Troubleshooting Trees and Instructions
7-7-10 Network Troubleshooting
7-7-10-1 No Connection to the Network at All
1.) Check that the network cable between the scanner and the wall network is connected and well
seated in both ends. (Use a network cable that is known to be OK.)
2.) Check the cable between the network-connector on the Back Panel to the LAN-connector
on the GEF.
3.) Connect a network cable between your Scanner and your PC. Try to ping from the Scanner to the
IP address on the PC. If OK, the hardware connection inside the Scanner is OK.