GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03)
D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
8-14 Section 8-7 - Software and Functional Checks after the Upgrade
Section 8-7
Software and Functional Checks after the Upgrade
1.) Press the UTILITIES key on the Touch Panel once to display the Utilities menu.
2.) Select SYSTEM SETUP
from the Utilities menu.
3.) Select
SYSTEM INFO to see which Software/Hardware version is installed in the unit.
4.) Verify the correct settings of the
OPTION page; see: Figure 8-2 on page 8-3.
If necessary, customize the settings according to the printout.
5.) Check and (if necessary) match the
NETWORK page settings according to the printout:
• “DICOM Configuration” dialog page; see: Figure 8-3 on page 8-4
• “Network Configuration” dialog page; see: Figure 8-4 on page 8-4
6.) Restart the system and perform basic functional checks to ensure that the system is functioning
normally.
Figure 8-20 System Setup - System Info page
Serial number
Application Software
Move the scroll bar downwards to review additional information about installed software/hardware.
EUM Version
Operating System
(e.g., MS-patch installed)
Backup File Name
System Type
Date of Factory Settings
Hardware ID
SLOT_CPU
(e.g., IBASE)