GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03)
D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-15
Section 8-8
Replacement or Activation of Options
Following Options are available:
•Real Time 4D
•DICOM
• RT_4D_Biopsy
•VOCAL
•XTD View
• DiagnoSTIC
• CRI (Compound Resolution Imaging)
•B-Flow
• VCI (Volume Contrast Imaging)
NOTE: Additional option fields are not yet implemented in the Voluson® 730Expert.
8-8-1 Operation for activating Options
1.) Touch the UTILITIES key on the Touch Panel once to display the Utilities menu.
2.) Touch the SYSTEM SETUP
key to invoke the setup desktop on the screen.
3.) Select the
OPTION page where you can see which options are installed in the system.
8-8-1-1 Operation for installing an Option:
1.) Position and select the option input field you want with the write cursor.
2.) Clear/Edit the current number, if one exists.
3.) Enter the encrypted serial code or option code with the keyboard.
4.) Click the SAVE&EXIT
button.
(The code will be checked and if correct the serial number/option is accepted.)
NOTE: After installation of the DICOM option, restart (turn of and on) the Voluson® 730Expert system.
Figure 8-21 Option page in the System Setup