GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03)
D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
8-22 Section 8-11 - Replacement of the Probe Holder
Section 8-11
Replacement of the Probe Holder
8-11-1 Replacement of the Probe Holder (at GW140)
8-11-1-1 Manpower
One person, 1 min.
8-11-1-2 Probe Holder - Replacement Procedure
1.) Check systems trolley version (refer to Table 9-3 on page 9-5). It has to be a GW140.
2.) Remove the probe holder from the bottom housing of the User Interface.
3.) Place the new probe holder on the bottom housing.
NOTICE
Replacement of the probe holder depends on the used trolley (GW).
GW130, not to be replaced by the user
GW140, see: Section 8-11-1 on page 8-22