GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03)
D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-23
Section 8-12
Replacement of the Trackball top fixation ring
8-12-1 Manpower
One person, 5 min.
8-12-2 Trackball top fixation ring - Replacement Procedure
1.) Remove the fixation ring by turning it counterclockwise.
2.) Mount the fixation ring by turning it clockwise.
Section 8-13
Replacement of Digipots and TGC Slider controls
8-13-1 Manpower
One person, 5 min.
8-13-2 Tools
small-sized slotted screwdriver or tweezers
8-13-3 Cap and/or Spring - Replacement Procedure
1.) Remove the cap (for Rotation digipots or for Slider-potentiometer TGC).
2.) Mount the cap (for Rotation digipots or for Slider-potentiometer TGC).
Figure 8-29 Trackball with top fixation ring
rotate counterclockwise
to remove the fixing ring
rotate clockwise to
mount the fixing ring