GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03)
D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
10-8 Section 10-5 - System Maintenance
10-5-2-2 Peripheral/Option Checks
If any peripherals or options are not part of the system configuration, the check can be omitted.
Refer to Table 3-7, “Approved Peripherals,” on page 3-43 for a list of approved peripherals.
10-5-3 Input Power
10-5-3-1 Mains Cable Inspection
10-5-4 Cleaning
10-5-4-1 General Cleaning
Frequent and diligent cleaning of the Voluson® 730Expert ultrasound system reduces the risk of
spreading infection from person to person, and also helps to maintain a clean work environment.
Table 10-6 Approved Peripheral/Hardware Option Functional Checks
Step Item Description
1VCR
Verify record/playback capabilities of the VCR. Clean heads and covers if necessary.
2 B/W Printer
Verify hardcopy output of the B/W video page printer. Clean heads and covers if necessary.
3 Color Printer
Verify hardcopy output of the Color video page printer. Clean heads and covers if necessary.
4 Line Printer
Verify hardcopy output of the Line printer. Clean heads and covers if necessary.
5 DICOM
Verify that DICOM is functioning properly. Send an image to a DICOM device.
6 Footswitch
Verify that the footswitch is functioning as programed. Clean as necessary.
7
DVD and/or
MO-Drive
Verify that the DVD-drive and/or optional MO-drive reads/writes properly
(export/recall image in Sonoview)
8 Modem
Verify modem remote connection
9ECG
Verify basic operation with customer.
Table 10-7 Mains Cable Inspection
Step Item Description
1 Unplug Cord
Disconnect the mains cable from the wall and system.
2 Inspect
Inspect it and its connectors for damage of any kind.
3Verify
Verify that the LINE, NEUTRAL and GROUND wires are properly attached to the terminals, and that
no strands may cause a short circuit.
4 Verify
Inlet connector retainer is functional.
Table 10-8 General Cleaning
Step Item Description
1 Console
Use a fluid detergent in warm water on a soft, damp cloth to carefully wipe the entire system.
Be careful not to get the cloth too wet so that moisture does not enter the console.
Caution: DO NOT allow any liquid to drip or seep into the system.
2
Monitor and
Touch Panel
Clean Top surface, Monitor and Touch Panel with a fluid detergent in warm water on a soft, damp cloth.
Caution: DO NOT spray any liquid directly onto the Covers, Monitor, keyboard, Touch Panel etc.