GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03)
D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-2 Section 3-2 - Set Up Reminders
3-2-2 Installation Warnings
1.) Since the Voluson® 730Expert weighs approximately 136 kg (300 lbs.) without options, two people
are required to unpack it. Two people are also required for installing any additional items in excess
of 16 kg / 35 pounds.(e.g., Monitor).
2.) There are no operator serviceable components. To prevent shock, do not remove any covers or
panels. Should problems or malfunctions occur, unplug the power cord.
Only qualified service personnel should carry out servicing and troubleshooting.
3-2-2-1 Moving/Lifting the System
3-2-2-2 System Acclimation Time
After being transported, the Voluson® 730Expert system may be very cold or hot.
It requires one hour for each 2.5° C increment it's temperature is below 10°C or above 40°C.
3-2-2-3 Control Panel Position
If weight is placed on the Control Panel (UI) in it’s extended position the console could tip over.
Figure 3-1 moving or lifting the system
CAUTION
!! CAUTION:
Equipment damage possibility. Turning the system on without acclimation after arriving at site
may cause the system to be damaged.
Table 3-3 Acclimation Time
°C
60 55 50 45 40 35 30 25 20 15 10 5 0 -5 -10 -15 -20 -25 -30 -35 -40
°F
140 131 122 113 104 96 86 77 68 59 50 41 32 23 14 5 -4 -13 -22 -31 -40
hrs
864200000002468101214161820
WARNINGWARNING
!! WARNING:
The system should NOT be moved with the Control Panel (UI) extended. Move the
Control Panel to it’s centered and locked position. Refer to Section 6-4 on page 6-5.
WARNINGWARNING
!! WARNING:
Monitor mounting mechanism may break if not properly supported (e.g., with packing
foam) during transportation.
When pulling, moving or lifting the system,
grasp it only at the rear handle of the trolley and
the handle underneath the foot rest.
WARNING
WARNING
Do NOT pull or lift the system with the front handle
of the user interface (operator panel).