GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03)
D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-8 Section 3-5 - Connection of Auxiliary Devices
Section 3-5
Connection of Auxiliary Devices
3-5-1 Preparations
NOTE: Normally the auxiliary devices and peripherals come already installed with the system.
1.) Carefully remove plastic caps using a knife and loosen screws and washers (see: Figure 3-5.)
2.) When the cover is loose on top, pull the rear cover out and move upwards.
3.) Disconnect the GND ground-cable from the back of the rear cover plate.
4.) Connect Peripherals according to correct connection scheme described in the following sub-
sections (refer to Table 3-6 on page 3-9).
5.) Connect the GND ground-cable at the rear cover plate.
6.) Mount rear cover plate, reattach screw and washers and plug-in the caps.
Figure 3-5 Remove rear cover plate
Figure 3-6 GND ground-cable
plastic caps; screws and washers
GND ground cable