GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03)
D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
Chapter 3 - Setup Instructions 3-9
3-5-1 Preparations (cont’d)
Table 3-6 below outlines Voluson® 730Expert hardware installation procedures described in the sub-
sections.
Table 3-6 Connection Procedures
Sub-section Description Page Number
3-5-2
Monitor Connection
3-10
3-5-3
Foot Switch Connection
3-10
3-5-4
ECG-preamplifier Connection
3-11
3-5-5
Global Modem Connection
3-12
3-5-6
S-VHS Video Recorder Connection
3-13
3-5-7
B/W Video Printer Connection
3-15
3-5-8
Line Printer Connection
3-16
3-5-9
Digital Color Printer Connection
3-17
3-5-10
Bluetooth Printer Connection
3-18
3-5-11
External USB-Devices
3-20
WARNINGWARNING
!! WARNING:
After each installation, the leakage currents have to be measured according to
IEC 60601-1 respectively UL 60601-1.
CAUTION
!! CAUTION:
Please observe that some printers may not be medical devices! If the Bluetooth Printer and/or
Line Printers are not medical devices, they have to be located outside of the patient
environment (according to IEC 60601-1 / UL 60601-1).
2,5 m
1,5 m
1,5 m1,5 m