GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03)
D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-20 Section 3-5 - Connection of Auxiliary Devices
3-5-11 External USB-Devices
3-5-11-1 External USB-Devices - Connection
When an external USB-storage device (such as a USB-memory stick or external hard disk) is connected
to the Voluson® 730Expert system, Windows detects the device and automatically installs a driver.
During this process, several dialogs may pop up, starting with the „Found New Hardware“ dialog.
The device is then accessible using the drive letter the system assigned to it.
3-5-11-2 External USB-Devices - Disconnection
Before an external USB-device (e.g., USB-memory stick) can be disconnected, the system has to be
informed about the removal of the device! For this purpose the System Setup -
BACKUP page (see:
Figure 4-22 on page 4-29) has a STOP USB DEVICES
button.
By clicking the STOP USB DEVICES
button, the „Unplug or Eject Hardware“ dialog is started.
Using this dialog, the USB-devices can be stopped before they are physically disconnected.
WARNINGWARNING
!! WARNING:
Do not connect or disconnect any external USB-devices to or from the system while
scanning a patient! The appearing dialogs could distract you from the scan!
Figure 3-18 Found New Hardware - USB Device
NOTICE
!! NOTICE:
When connecting external USB devices, be sure to execute Safety Directions found in Chapter 2 of the
Voluson® 730Expert Basic User Manual.
CAUTION
!! CAUTION:
Unplugging or ejecting USB devices without first stopping them can often cause the system to
crash and possibly result in loss of valuable data.
Figure 3-19 Unplug or Eject Hardware