GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03)
D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
Chapter 3 - Setup Instructions 3-25
3-6-4 Transducer Connection
Connect a transducer to one of the three rightmost transducer receptacle as follows:
1.) Inspect the probe and probe socket to verify that it is free of debris.
2.) Ensure that the transducer twist lock lever is at horizontal position.
3.) Insert the connector on the receptacle guide pin until it touches the receptacle mating surface.
4.) Twist the transducer twist lock lever to vertical position to lock it in place.
Twist the lever to the horizontal position to disconnect the transducer.
5.) Open the right-hand side door, lay the cable into the intended cable holder and close the door.
NOTE: Prior to connecting or disconnecting a probe, freeze the image.
It is not necessary to turn OFF power to connect or disconnect a transducer.
Figure 3-25 Transducer Connection
CAUTION
!! CAUTION:
If the cable spout on the right-hand door is missing, don’t pull the probe cable.
Please insert the spout in the designated place to avoid damage of the probe cable.
CAUTION
!! CAUTION:
Do not bend the probe cable acutely. Fault conditions can result in electric shock hazard.
Do not touch the surface of probe connectors which are exposed when the probe is removed.
Do not touch the patient when connecting or disconnecting a probe.