GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03)
D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
Chapter 3 - Setup Instructions 3-43
3-8-2 On-Board Optional Peripherals
Mains outlets: Mains socket ST1, ST2, ST3, ST4, ST5 for accessories.
All mains outlets are co-switched by the unit’s mains switch via built-in isolation transformer.
Output voltage for: ST1 - ST5: 115V or 230V.
CAUTION
!! CAUTION:
Modification of voltage setting only by an authorized service person!
The maximum power consumption of equipment (inclusive color video monitor) connected to
these outlets must not exceed 350VA!
Table 3-7 Approved Peripherals
Device Manufacturer Model Video Signal
B/W Video Printer SONY UP-895MD NTSC/PAL
Digital Color Video Printer
SONY UP-D21MD USB-Port
SONY UP-D23MD USB-Port
Video Cassette Recorder
SONY
SVO-9500MD
SVO-9500-MDP
NTSC
PAL
Mitsubishi
HS-MD3000U
HS-MD3000E
NTSC
PAL
Line Printer
Hewlett Packard hp deskjet 990cxi USB-Port
Hewlett Packard hp deskjet 995c USB-Port
Color Deskjet Printer (Bluetooth)
Hewlett Packard HP 5600 Series USB-Port (Bluetooth)
Olivetti Olivetti Job Jet 210 USB-Port (Bluetooth)
Hewlett Packard HP 5940 USB-Port (Bluetooth)
Canon Pixma MP600 / MP610 USB-Port (Bluetooth)
Magneto-Optical Drive (MOD) Fujitsu MCM3130AP
Global Modem Multi-Tech MT5634ZBA
ECG Preamplifier MAN6
Footswitch MFT7