Intended use:
This equipment is to provide a laboratory test within the range of
1000 4500 rpm for test tubes containing in vitro diagnostic
specimens for qualitative or quantitative test procedures. Refer to the
clinic laboratory method specified by the reagent manufacturer or
established by medical technology for products applications.
Installation:
1
Strip the packing materials
2
Open the cover of the carton box and remove the artificial form
3
Carry the centrifuge carefully from the carton box;
4
Remove the plastic bag from the centrifuge;
5
Place centrifuge onto a sturdy and flat table.
6
Don’t drop the equipment and don’t place the equipment
upside down.
Content
- Rotor : 1
- Carbon brush : 1 pair spare carbon brush
- Fuse : 1
- Adaptor : Corresponding to the rotor.
- Power Cord
3-wire
CSA type, SVT 105ºC, 18AWG x 2 C
Definition
g Force
The measurement for samples undergoing the stress of
acceleration in a centrifuge.
RCF
Relative Centrifugal Force
RCF is the centrifugal force that the samples undergoes which
relies on the speed of rotation
N
in rpm and the rotating radius
(cm)
R
. The rotating radius is measured from the center axis of
rotor to the extreme end of the centrifuge tubes.
Formula: RCF = 11.18 x R x ( N / 1000)²
A-0815: Ø 21.1cm;
A-1215: Ø 22.7cm