TABLE TOP CENTRIFUGE
PLC-01,PLC-02,PLC-03,PLC-04,PLC-05
USER’S MANUAL
OPERATION / MAINTENANCE
AND PARTS LIST
! Safety Information
This manual contains important and safety information.
The user must carefully read and understand the contents of this
manual prior to using this equipment.
2013.0115.2000

Questions and Answers

  • B
    Brandi WilsonAug 5, 2025
    Why is my Gemmy PLC-05 vibrating?
    • A
      andrewnelsonAug 5, 2025
      Vibration in your Gemmy Laboratory Equipment can occur due to a rotor imbalance. Make sure to place tubes symmetrically. Another possible cause is a defective rubber support; in this case, replace it with a new one.
  • F
    fandersonAug 9, 2025
    What to do if Gemmy Laboratory Equipment failure of operation?
    • H
      henrychrisAug 9, 2025
      If your Gemmy Laboratory Equipment fails to operate, it could be due to several reasons. Worn-out brushes might be the cause, so replace the carbon brushes. Alternatively, the motor or PCB could be defective, requiring replacement of those parts. Also check for a blown fuse and replace it if necessary. Finally, ensure the linkage is connected; if disengaged, connect it.
  • M
    Mark GuerraAug 13, 2025
    Why is my Gemmy PLC-05 speed erratic?
    • Y
      yatkinsonAug 13, 2025
      Erratic speed in your Gemmy Laboratory Equipment may be caused by worn-out carbon brushes; replace them with new ones. Another possibility is a defective triac or VR, which would require replacing those parts.
  • J
    James DavisAug 17, 2025
    What to do if Gemmy PLC-05 Laboratory Equipment motor spinning failure?
    • C
      Corey PhillipsAug 18, 2025
      If the motor in your Gemmy Laboratory Equipment fails to spin, several factors could be responsible. First, check the carbon brushes for wear and replace them if needed. Second, the PCB might be malfunctioning, requiring replacement. Third, the motor itself could be burnt out, necessitating replacement. Finally, ensure all wiring is properly connected; reconnect any disengaged wiring.
  • J
    julia28Aug 21, 2025
    Why is rotor tilted in my Gemmy PLC-05?
    • G
      Gene RileyAug 22, 2025
      A tilted rotor in your Gemmy Laboratory Equipment can be due to a loose fastener attached to the motor; fix and glue the fastener. It can also happen if the rubber support is broken, so replace it with a new one.
  • H
    Heather BrownAug 26, 2025
    How to fix change of MAX and MIN rpm in Gemmy PLC-05 Laboratory Equipment?
    • P
      Pamela MooreAug 27, 2025
      If you notice a change in the MAX and MIN RPM of your Gemmy Laboratory Equipment, it may be due to a loose SVR on the PCB. Calibrate the speed. See Page 7 for correction
  • J
    Joseph BrooksAug 30, 2025
    What to do if timer can’t work on Gemmy PLC-05?
    • M
      Margaret MayAug 31, 2025
      If the timer on your Gemmy Laboratory Equipment isn’t working due to a counting failure, turn the timer to 60 minutes, then turn it to 0 before setting the desired time.

Summary

Safety Information

Equipment Classification and Standards

Classification as Class I equipment and compliance with safety standards.

Intended Use, Installation, and Definitions

Tubes and Grounding

Grounding Instructions

Correct connection of the grounding wire and warnings.

Operation Procedures

Centrifuge Operation

Steps for operating the centrifuge, setting time, and speed.

Calibration and Performance

Recalibration Procedure for Speed

Steps to recalibrate the centrifuge's min/max speed settings.

Maintenance and Troubleshooting

Maintenance and Servicing Guidelines

Procedures for cleaning, inspection, and servicing components.

Troubleshooting Common Problems

Table of problems, possible causes, and corrective actions.

Gemmy PLC-05 Specifications

General IconGeneral
Input Voltage24V DC
Number of Inputs8
Communication InterfaceRS-485
TypeProgrammable Logic Controller
Storage Temperature-20 to 70°C
Humidity5% to 95% (non-condensing)