OPERATION:
- Check the specification label located on back of this equipment.
- Plug cord in a properly grounded outlet.
- Place the test tubes into the tube adaptors. Be sure to place the
tubes diagonally and symmetrically to keep the rotor balanced.
- Close the cover completely. NOTE, Locking latch is optional.
- Dial preset your desired time by turning the control knob for any
spin interval from 0 60 minutes. This equipment is working by
timer-activated operation.
- When turn clockwise the timer, the pilot lamp will light up at the
same time, which indicates the power up.
- Adjust the control knob of speed regulator to bring speed to your
desired operation speed and centrifugal force.
- This centrifuge does not include the speed indicator, the reference
of speed and RCF is depicted below. Nevertheless, the user is
recommended to measure the speed by a hand held LCD
tachometer in order to achieve a correct speed and RCF.
- Speed and RCF varies slightly centrifuge to centrifuge.
- The bell rings and power will shut out automatically when the set
time has elapsed and the pilot lamp will distinguish simultaneously.
WARNING
- Do not open the lid only until the rotor has stopped completely.
- Do not stop the rotor by hand. There is possibility of personal injury.
- Gemmy will not ensure the responsibility if user violate the above
regulation.