765DC
15
Intraoral X-Ray Tube Housing
Product Data Sheet
Gendex
©
765DC
The Gendex
©
765DC housing is designed to operate with the Gendex
©
GX 70-10DC insert. The
housing has been designed to operate on the Gendex
©
765DC high frequency high voltage
control, making up the Gendex
©
765DC Intraoral Dental X-Ray system.
Specifications
Shockproof Housing: Zinc, hermetically sealed under vacuum with high
dielectric insulating oil. Inherent to the housing is an
expansion chamber designed to provide adequate
compensation for the full temperature range.
Operating Temperature: +50°F to 100°F ≈ (+10˚C to +40˚C)
Storage Temperature: -40°F to 160°F ≈ (-40˚C to +70˚C)
Inherent Filtration:
Radiation Output:
Technique 65kV, 7mA, 0.50sec. Source to Detector Distance 230mm (9”)
410mR ±125mR (3.55 mGy ± 1.05 mGy)
Typical Half Layer: 1.9mm Aluminum at 65kV
Thickness / mm.
AHE
1
/mm.
Al. Equivalent
Material Min. Max. 65 kV Min. Max.
Glass 1.27 1.91 0.51 0.65 0.97
Insulating Oil 6.5 7.7 0.08 0.52 0.62
Polycarbonate 16.4 16.7 0.09 1.48 1.50
Total 2.64 3.09