Leakage Radiation: Complies to 601-1-3 IEC Section 29.204 “Leakage
Radiation”. Maximum continuous power input equals
1,400 heat units per min. (HU/Min.) tube is the limiting
factor. HU/Min. = (kV) x (mA) x (1.4) x (Exp. Time
(sec) / 60)
1,400 HU/Min. = 65 kV x 0.27mA x 1.4 x 60 sec.
X-Ray Coverage: The insert target angle is 12.5°
• Circular cone at 8 inches (200 mm) SSD
2
60 mm
Dia. +0, -3mm.
• Circular cone at 12 inches (300 mm) SSD
2
60 mm
Dia. +0, -3mm.
• Rectangular cone at 8 inches (200 mm) SSD
2
36mm
x 46mm
1
AHE Aluminum Half value Equivalent correlation
2
SSD Source to Skin Distance
Weight: 13.0 lbs. 5.9 kg.
Maximum Head Voltage: Anode to Cathode 65 kV
Anode to Ground 70 kV
Thermal Characteristics:
• Housing Heat Storage Capacity 250 kHU’s (177J)
(see Graph for more detail)
• Maximum Cooling Rate 3.3 kHU’s/Min.
Heat Units = (kV * mA * time in seconds * 1.4)
Gendex 765DC Housing
Heating & Coling Chart
20
25
30
35
40
45
0
2
5
50
75
1
0
0
125
1
5
0
1
75
200
2
25
250
275
3
0
0
3
25
350
Time (min.)
Temperature (°C)
0
50
100
150
200
250
300
Heat Units (kHu's)
Heating, 100% Duty Cycle, ambient