the sleeve O-ring over the new sleeve to secure it. Reach inside of the glovebox with the
other glove and pull the old sleeve off and into the work area. Dispose of Sleeve through
the airlock.
Glove and Sleeve Inspections
A visual inspection of all glove and sleeve surfaces is recommended at the beginning and end
of each day. Any tears or punctures would require a replacement.
If the unit is operating and is under pressure the operator can run a hand on all surfaces to
feel for any air moving into or out of the work area through imperfections in the surfaces.
Soap solution can also be used. If the LFGI has pressure the soap will bubble at any cut,
tear or puncture.