Sharps
The Sharps disposal system on the LFGI is
designed to accommodate any sharps
container with a 5” mouth or hinged lid up to
a maximum size of 27” H x 20 “ D x 14.5” W
Some BD part numbers that fit this
description are as follows:
305491
300473
305615
305616
305601
305602
305609
305610
305665
305666
305457
305493 * Yellow for Chemo
305603 * Yellow for Chemo
305614 * Yellow for Chemo
The Sharps Collector shelf is height adjustable and should be placed at a height where the
sharps tube is down inside the collector. This will reduce the risk of overflowing a collector.
All institute regulated safety policies should be followed when handling sharps.
To move the shelf simply tilt up the outside edge of the shelf to release the bottom tabs
from the slotted track. The top shelf tabs are notched and will require a push up from the
bottom of the shelf. This system is equivalent to the casework found in many pharmacies
and allows for many different sizes of containers to be used.
Alternatively, sharps collectors can be placed inside of the work area and can be passed
through the airlock.
Waste Removal
Waste discharge tubes are used for the rapid removal of waste products generated during
procedures. The Discharge Tubes need to be kept sealed with the caps on while not in use.