Introduction
The Stainless Steel Laminar Flow Glovebox /Isolator is a complete barrier system. It
provides sterile laminar flow air for aseptic pharmacy preparations while protecting
pharmacy personnel from hazardous materials.
The Laminar Flow Glovebox /Isolator uses an amalgamation of cleanroom and
containment technologies designed specifically for critical pharmacy applications. This unit
provides laminar flow air for product protection and complete containment for operator
protection.
The Laminar Flow Glovebox /Isolator offers the highest level of product protection by
providing vertical unidirectional (laminar) flow HEPA filtered air to the complete work
environment. The unit utilizes a full width and depth supply HEPA filter above the work
surface. The entire work area is bathed in HEPA filtered unidirectional mass displaced air.
This provides a better than ISO Class 5 particulate-free environment. Particulates generated
by manipulations are continuously removed, maintaining ISO Class 5 or better conditions
when operational. All air entering the LFGI is twice filtered, once by the pre-filter then by
the supply HEPA filter.
8