Unscrew the four unmarked screws on the underside of the case cover
with a 4 mm key.
Remove the case cover (1).
Unplug connector (2) by pulling it upwards as shown by the arrow.
Using a screwdriver (3):
turn screw (4) clockwise – to raise the temperature -
turn screw (4) anticlockwise – to lower the temperature -.
Refit the case cover (1).
Danger of electric shock
See Section 2.3, page 3.
Contamination of the sterilised goods
“Danger of patient infection”
See Section 2.5, page 4, item 1.