2001/2 / 2001/4 Wasserdestillierapparate Water Stills
Seite 12 Rev.04 vom / dated 01.04.2020
3. Before Installation
Important information are marked in bold letters in these instructions,
safety indications are additionally marked by the following warning
symbols and mandatory signs.
Warning of hot
liquids and
vapour
Warning of
hot surfaces
Read and observe the
operating instructions
Warning of
dangerous
electrical
voltage
Before maintenance and repair
disconnect the unit all-pole
from the electrical mains (pull
the plug from the socket).
All free openings of the water connections on the right-hand side of the unit are taped close.
These transport protections have to be removed before installation.
4. Set-up and Location of the Water Still
Place on solid, even and level surfaces inside buildings only. Make sure to place the unit
only on a watertight, temperature-resistant, non-flammable surface. The location must
provide sufficient space as well as the necessary carrying capacity for the unit (unit weight as
per technical data, chapter 11 of this manual, plus weight of the filling.
The unit is not suitable for use in potentially explosive areas.
5. Operating Voltage
The Water Still must be connected to the mains through a correctly
installed shock-proof socket or through an on-site main switch. The Still
is a protection class I electrical appliance, a connection to the earth
conductor (PE) must be ensured. For information on the required
mains fuse please view Technical Data, chapter 11 of this manual.
The electrical connection must ensure an all-pole separation of the
Water Still from the mains at any time.
The mains connection cable must not touch any hot surfaces of the
unit. It may not lead underneath the unit.
The main switch of the unit must be off (position O). The voltage on the
nameplate (at the back of the unit) must be identical to the mains
voltage. If they are identical, connect the unit to the mains.
For further information, please also refer to chapter 13 of these operating
instructions "Connection to the Mains Supply".
6. Water Connections
All water connections of the
Water Still, with the exception
of the distillate withdrawal at
the front, are situated at the
right-hand side of the unit.
Hoses for water inlet and
outlet are not included in the
standard scope of supply.