2001/2 / 2001/4 Wasserdestillierapparate Water Stills
Seite 17 Rev.04 vom / dated 01.04.2020
11. Technical Data
Exterior dimensions (W x D x H) 270 x 250 x 510 mm
Distillation capacity Model 2001/2 2 l / h distillate
Model 2001/4 4 l / h distillate
Distillate quality Mono distillate
approx. 2.3 S/cm at 2C
The conductivity of the distilled water is directly
related to the chemical composition of the feed
water. Components of the feed water having the
same of a lower evaporation point than water may
deteriorate the conductivity.
Cooling water requirement Model 2001/2 20 litres / hour
Model 2001/4 40 litres / hour
Low water cut-off (thermostat) Electromechanical temperature
limiter, with capillary tube sensor.
Switch-off temperature 135°C / -15K.
Required cooling water pressure 1-2 bar
14.5 psi / 29.0 psi
Electrical connection /
Mains connection
Caution!
Mains voltage deviations, even within the mentioned
limits, influence the quantity of distillate produced!
Model 2001/2
Model 2001/4
230V +/-10%,
50/60 Hz, 2.0 kW
Schuko plug CEE 7/7
230V +/-10%,
50/60 Hz, 3.0 kW
Schuko plug CEE 7/7
Mains fuse On-site Model 2001/2 10 Amp. (max. 16A)
Internal Model 2001/2 10 Amp. inert
On-site Model 2001/4 16 Amp.
Internal Model 2001/4 15 Amp. inert
Protection class / - type I / IP20
Surrounding conditions Use only inside buildings
(not in potentially explosive areas)
Up to 2000 m MSL
+10 °C to +40 °C
max. 80% rel. humidity to 31°C,
decreasing to 40% rel. humidity
at 40°C
Weight
Net / with water filling
Model 2001/2
Model 2001/4
7.7 kg / 12.2 kg
8,0 kg / 12.5 kg