MaxiMet Page 142 Issue 3
Doc. No. 1957-PS-021 May 2018
GPSSTATUS - Location Fix and Number of Satellites
Result e.g. 010B.
Where: 0 is padding.
1 is GPS SPS mode fix valid (0 is fix not available).
0B is a hexadecimal representation of the number of satellites acquired, 11 satellites found. 0A
would be 10 satellites etc.
TIME (and Date) - MaxiMet Date and Time (can be updated by GPS
Option)
Time can be set manually by the user or updated by the GPS Module (if fitted). GPS time can be
automatically updated every hour (GPS default setting).
By default time will be UTC.
Result e.g. 2015-06-04T10:01:36.8.
VOLT - Supply Voltage
DC Supply voltage measured at the MaxiMet.
Result e.g. +10.5 (volts).
STATUS - Status of MaxiMet Sensors
Outputs the MaxiMet Sensors Status Code.
Result e.g. 0000 (see para 8.3).
WINDSTAT - Status of Wind Sensor in detail
Outputs status codes relating to wind sensor data.
Result e.g. 0000 (see para 8.3).
USERINF - User Information Field
The User Information field shall have a maximum size of 16 characters, comprised of printable ASCII
characters (to exclude "@", ",", "*", "%" and "?").
Note: Space character may be included.
This field allows for instance to record installation or calibration dates.
Result e.g. CAL DUE 15.08.18