NOTICE
Do not lubricate this pipette.
The use of lubricant cancels
the warranty of this pipette.

Other manuals for Gilson PIPETMAN P1000

Questions and Answers

  • E
    emily18Aug 4, 2025
    Why do tips fall off my Gilson PIPETMAN P1000?
    • W
      William StewartAug 4, 2025
      If the tips on your Gilson Laboratory Equipment pipette are falling off or not fitting correctly, several factors could be at play. It might be due to using low-quality tips, in which case you should replace the tip. Alternatively, the tip holder or tip ejector could be damaged. If the tip holder is damaged, replace the tip. If the tip ejector is damaged, ensure the ejector spacer is correctly placed on the tip ejector. Lastly, check if the ejector lock is misaligned; if so, align it properly.
  • E
    Eric ShawAug 9, 2025
    What to do if tips fall off Gilson Laboratory Equipment?
    • C
      Cynthia MorrisAug 9, 2025
      If the tips fall off or do not fit correctly on your Gilson Laboratory Equipment pipette, it might be due to low-quality tips. Try using Gilson DIAMOND tips. Another reason could be damaged tip-holders, in which case you should replace the tip-holder. Also, a damaged tip-ejector can cause this issue, so consider replacing the tip-ejector.
  • C
    Carol OlsonAug 13, 2025
    What to do if Gilson PIPETMAN P1000 Laboratory Equipment pipette is leaking?
    • J
      Jennifer DavisAug 13, 2025
      If your Gilson Laboratory Equipment pipette is leaking sample, it could be due to a damaged tip holder, in which case you should replace both the tip holder and the tip. Another potential cause is worn O-ring or seal parts, and the solution is to replace the tip.
  • S
    stephanie65Aug 17, 2025
    How to fix a Gilson PIPETMAN P1000 Laboratory Equipment pipette that won't aspirate?
    • P
      Paul MunozAug 17, 2025
      If your Gilson Laboratory Equipment pipette won't aspirate, it could be due to a damaged or corroded piston. In this case, return the pipette to the supplier. Also, it might be due to improper repair. Refer to assembly page 15.
  • S
    Sheri OdonnellAug 21, 2025
    Why is my Gilson PIPETMAN P1000 pipette leaking?
    • E
      erikevansAug 21, 2025
      If your Gilson Laboratory Equipment pipette is leaking sample, it could be due to a damaged tip-holder, which requires replacing the tip-holder. Alternatively, the O-ring or seal might be worn, in which case you should replace both parts.
  • J
    john05Aug 26, 2025
    Why is Gilson PIPETMAN P1000 Laboratory Equipment pipette not precise?
    • M
      melaniewernerAug 27, 2025
      If your Gilson Laboratory Equipment pipette is not precise, it could be due to several reasons. The tip-holder or connecting nut may be loose; in this case, tighten the connecting nut. It may also be due to incorrect operator technique, and operator training is needed. Damaged or corroded pistons can also cause this, so return the pipette to the supplier. Also, damaged tip-holders can cause this, so replace the tip-holder. Finally, a worn O-ring or seal may be the cause, so replace both parts.
  • S
    stephenschristineAug 30, 2025
    What to do if Gilson PIPETMAN P1000 Laboratory Equipment pipette won’t aspirate?
    • J
      Juan HallAug 30, 2025
      If your Gilson Laboratory Equipment pipette won’t aspirate, it could be due to a worn O-ring (replace both parts), a damaged tip-holder (replace the tip-holder), or a loose connecting nut (tighten the connecting nut). A damaged or corroded piston could also be the cause, requiring you to return the pipette to the supplier. Improper repair or assembly could also cause this problem.

Summary

Introduction to PIPETMAN Classic

Parts Checklist

Verifying included items after unpacking the pipette to ensure completeness and condition.

Setting the Volume

Pipetting Procedures

Fitting Pipette Tips

Instructions on how to properly attach PIPETMAN DIAMOND tips for an airtight seal.

Aspirating Liquid

Step-by-step instructions for drawing the correct volume of liquid into the pipette.

Dispensing Liquid

Step-by-step instructions for expelling the aspirated liquid from the pipette.

Guidelines for Good Pipetting

Best practices and techniques for achieving accurate and precise pipetting.

Troubleshooting Common Issues

Troubleshooting Table

A guide to identifying pipette symptoms, possible causes, and recommended actions.

Pipette Inaccuracy Issues

Diagnosing and resolving problems leading to inaccurate volume measurements.

Pipette Imprecision Issues

Diagnosing and resolving problems affecting pipette precision.

Leak Test Procedure

A test to verify that the pipette does not leak liquid.

Maintenance and Care

Cleaning and Decontamination

Methods for cleaning and decontaminating pipette parts for hygiene and longevity.

Spare Parts Information

Warranty Information

Gilson PIPETMAN P1000 Specifications

General IconGeneral
Volume Range100 - 1000 µL
Increment10 µL
Tip EjectorYes
Adjustable VolumeYes
Length220 mm
AutoclavableYes