Gilson PIPETMAN P20 - User Manual

Gilson PIPETMAN P20
15 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
Users Guide
EN
CLASSIC
TM
NOTICE
Do not lubricate this pipette.
The use of lubricant cancels
the warranty of this pipette.

Other manuals for Gilson PIPETMAN P20

Questions and Answers

  • R
    ricescottAug 7, 2025
    Why do tips fall off my Gilson Laboratory Equipment?
    • M
      Michelle RoseAug 7, 2025
      If the tips on your Gilson Laboratory Equipment pipette are falling off or not fitting correctly, several factors could be at play. It might be due to using low-quality tips, in which case you should replace the tip. Alternatively, the tip holder or tip ejector could be damaged. If the tip holder is damaged, replace the tip. If the tip ejector is damaged, ensure the ejector spacer is correctly placed on the tip ejector. Lastly, check if the ejector lock is misaligned; if so, align it properly.
  • J
    Jeffrey HowardAug 11, 2025
    What to do if tips fall off Gilson Laboratory Equipment?
    • T
      Travis CohenAug 12, 2025
      If the tips fall off or do not fit correctly on your Gilson Laboratory Equipment pipette, it might be due to low-quality tips. Try using Gilson DIAMOND tips. Another reason could be damaged tip-holders, in which case you should replace the tip-holder. Also, a damaged tip-ejector can cause this issue, so consider replacing the tip-ejector.
  • S
    Steven MartinAug 16, 2025
    What to do if Gilson PIPETMAN P20 pipette is leaking?
    • K
      Kimberly NelsonAug 16, 2025
      If your Gilson Laboratory Equipment pipette is leaking sample, it could be due to a damaged tip holder, in which case you should replace both the tip holder and the tip. Another potential cause is worn O-ring or seal parts, and the solution is to replace the tip.
  • V
    Veronica GrahamAug 20, 2025
    How to fix a Gilson Laboratory Equipment pipette that won't aspirate?
    • D
      David ButlerAug 20, 2025
      If your Gilson Laboratory Equipment pipette won't aspirate, it could be due to a damaged or corroded piston. In this case, return the pipette to the supplier. Also, it might be due to improper repair. Refer to assembly page 15.
  • A
    Ashley LeeAug 25, 2025
    Why is my Gilson PIPETMAN P20 pipette leaking?
    • S
      Sarah GonzalesAug 26, 2025
      If your Gilson Laboratory Equipment pipette is leaking sample, it could be due to a damaged tip-holder, which requires replacing the tip-holder. Alternatively, the O-ring or seal might be worn, in which case you should replace both parts.
  • D
    Danielle HowardAug 28, 2025
    Why is Gilson PIPETMAN P20 pipette not precise?
    • H
      Heather YoungAug 28, 2025
      If your Gilson Laboratory Equipment pipette is not precise, it could be due to several reasons. The tip-holder or connecting nut may be loose; in this case, tighten the connecting nut. It may also be due to incorrect operator technique, and operator training is needed. Damaged or corroded pistons can also cause this, so return the pipette to the supplier. Also, damaged tip-holders can cause this, so replace the tip-holder. Finally, a worn O-ring or seal may be the cause, so replace both parts.
  • L
    Leslie RussellSep 2, 2025
    What to do if Gilson Laboratory Equipment pipette won’t aspirate?
    • M
      Megan MoralesSep 2, 2025
      If your Gilson Laboratory Equipment pipette won’t aspirate, it could be due to a worn O-ring (replace both parts), a damaged tip-holder (replace the tip-holder), or a loose connecting nut (tighten the connecting nut). A damaged or corroded piston could also be the cause, requiring you to return the pipette to the supplier. Improper repair or assembly could also cause this problem.

Gilson PIPETMAN P20 Specifications

General IconGeneral
BrandGilson
ModelPIPETMAN P20
CategoryLaboratory Equipment
LanguageEnglish