Guidance and manufacturer’s declaration – electromagnetic immunity
The Fingertip Oximeter is intended for use in the electromagnetic environ-
ment specied below. The customer or the user of The Fingertip Oximeter
should assure that it is used in such an electromagnetic environment.
Immunity
test
IEC60601
test level
Compliance
level
Electromagnetic environment -
guidance
Conducted
RF IEC
61000-4-6
Radiated
RF IEC
61000-4-3
3 Vrms
150 kHz
to 80 MHz
3 V/m
80 MHz to
2.5 GHz
N/A
3 V/m
Portable and mobile RF com-
munications equipment should
be used no closer to any part of
The Fingertip Oximeter, including
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
Recommended separation
distance
80MHz to 800MHz
800MHz to 2.5GHz
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the trans-
mitter manufacturer and d is the
recommended separation dis-
tance in metres (m).
b
Field strengths from xed RF
transmitters, as determined by
an electromagnetic site survey,
a
should be less than the com-
pliance level in each frequency
range.
b
Interference may occur in the vi-
cinity of equipment marked
with the following
symbol.
d= 1.2 √P
d= 1.2 √P
d= 2.3 √P