Guidance and manufacturer’s declaration-electromagnetic emission
The Fingertip Oximeter is intended for use in the electromagnetic environment
specied below. the customer or the user of the Fingertip Oximeter should
assure that it is used in such an environment.
Immunity test IEC60601
test level
Compliance
level
Electromagnetic
environment-guidance
Electrostatic
discharge
(ESD) IEC
61000-4-2
±6kV contact
±8kV air
±6kV
contact
±8kV air
Floors should be wood,
concrete or ceramic tile.
If oor are covered with
synthetic material, the
relative humidity should
be at least 30%.
Electrical fast
transient/
burst
IEC 61000-4-4
±2kV for power
supply lines
±1kV for input /
output lines
N/A N/A
Surge IEC
61000-4-5
±1kV line (s)
to line(s)
±2kV line(s)
to earth
N/A N/A
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
<5% UT
(>95% dip in UT)
for 0.5 cycle
<40% UT
(60% dip in UT)
for 5 cycles
<70% UT
(30% dip in UT)
for 25 cycles
<5% UT (>95%
dip in UT) for 5 s
N/A N/A
Power
frequency
(50/60Hz)
magnetic eld
IEC 61000-4-8
3A/m 3A/m Power frequency
magnetic elds should
be at levels characteris-
tic of a typical location
in a typical commercial
or hospital environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level.