Chapter 11 Appendix
11.1 Prompt information explanations
Mute C-D: XXX seconds
Alarm silence count down: XXX seconds
NIBP C-D: XXX seconds
NIBP auto measuring cycle count down: XXX seconds
TOUR C-D: XXX seconds
Hemostat alert count down: XXX seconds
Probe off
SpO
2
probe is off form the patient or disconnected from the device
PR over limit
PR value exceeds the high/low alarm limit
SpO
2
over limit
SpO
2
value exceeds the high/low alarm limit
SYS over limit
Systolic pressure value exceeds the high/low alarm limit
DIA over limit
Diastolic pressure value exceeds the high/low alarm limit
MAP over limit
MAP value exceeds the high/low alarm limit
NIBP error 1#
Sensor or other hardware error
NIBP error 2#
Very weak signal because of the cuff, or the patient has very weak pulse
NIBP error 3#
Blood pressure amplifier overflow due to excessive movement
NIBP error 4#
Leaking during the pneumatic device testing
Cuff error
Cuff is not wrapped correctly, or is not connected
NIBP error 5#
Hardware fault of NIBP module
Air leak
Pneumatic part, tube or the cuff leak air
NIBP over range
The measurement range exceeds 255mmHg (for neonates: over 135 mmHg)
Over motion
Excessive movement or noise during inflation and measurement. Another
measurement will be taken.
Over pressure
Cuff pressure exceeds the safety limit value of software. (limit for adults:
290mmHg; limit for pediatric: 145mmHg)
Or caused by cuff extrusion or flapping cuff with force.
NIBP timeout
Adult measurement is more than 120 seconds, neonate measurement is more
than 90 seconds.