GL820-UM-251-9370 5-12
5. INSPECTION AND CHECK PROCEDURES
(4) Press the [CH GROUP] key to display from the CH11 to the CH20.
(5) Press the [START/STOP] key to measure.
(6) Wait until the "Finished" message is displayed, and then display the values as shown below.
Message
Conrm that this value is within
the below ratings after data has
been captured.
(7) Check that the level of measurement is within below rating.
RTD precision ratings
Range Input resistor
Reference
Temperature
Factory Rating Product Rating
Pt100 350 Ω 715.3 ˚C +714.5 ˚C to +716.1 ˚C +714.3 ˚C to +716.3 ˚C
Pt1000 2500 Ω 408.8 ˚C +407.8 ˚C to +409.2 ˚C +408.0 ˚C to +409.6 ˚C