EasyManua.ls Logo

GRAPHTEC GL840 - Specification of Input Section (GL840-WV Withstand High Voltage High-Precision Terminal); Installation Guide

GRAPHTEC GL840
24 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Installation Guide
22
For the installation procedure of the GL840 application software (USB driver /
GL100_240_840-APS), refer to the “Application Software Manual” included in the
attached CD-ROM.
Specification of input section (GL840-WV Withstand High Voltage high-precision terminal)
Description
Item
Voltage : ± (0.05% of F.S. +10 µV)Measurement accuracy *
1
(23°C ±5°C)
• When 30 minutes or
more have elapsed after
power was switched on
• Sampling 1 s/10 ch
• Filter ON (10)
• GND connected
Between input terminal/GND : 300Vp-p
Between input terminal/input terminal : 1 minute at 600Vp-p
Between input terminal/GND : 2300 VACrms 1 minute
Between input terminal/input terminal : 600Vp-p
Between +/– terminals : 20mV to 2Vrange (60Vp-p)
5V to 100V range (110Vp-p)
Maximum input voltage
Withstand voltage
*1: Thermocouple diameters T, K: 0.32 , others: 0.65 
Resistance temperature detector
Type Measurement AccuracyMeasurement
Temperature Rang
R/S
B
K
E
T
N
W
Reference contact compensation
accuracy
J
0TS100°C ±4.5˚C
±3.0˚C
±2.2˚C
±2.2˚C
±2.5˚C
±1.5˚C
±0.8˚C
±1.0˚C
±0.8˚C
±1.5˚C
±0.6˚C
±0.6˚C
±2.2˚C
±1.0˚C
±1.8˚C
±3.5˚C
±1.0˚C
±0.8˚C
±0.3˚C
100<TS300°C
R: 300<TS1600°C
S: 300<TS1760°C
400TS600°C
600<TS1820°C
-200TS-100°C
-100<TS1370°C
-200TS-100°C
-100<TS800°C
-200TS-100°C
-100<TS400°C
-200TS-100°C
-100<TS100°C
100<TS1100°C
* 3-wire system
-200TS<0°C
0TS1300°C
0TS2000°C
Thermocouple
Type Applied current Accuracy
Measurement
Temperature Range
Pt100
JPt100
Pt1000
-200TS100°C ±0.6°C1mA
100<TS500°C ±0.8°C
1mA
500<TS850°C
-200TS100°C
100<TS500°C
-200TS100°C
100<TS500°C
±1.0°C
±0.6°C
±0.8°C
±0.6°C
±0.8°C
0.3mA

Other manuals for GRAPHTEC GL840