±7.0°C
±5.0°C
±(0.05% of rdg +3.0°C)
±(0.05% of rdg +3.0°C)
±5.5°C
±(0.05% of rdg +3.0°C)
±(0.05% of rdg +3.0°C)
±(0.05% of rdg +2.0°C)
±(0.05% of rdg +3.0°C)
±(0.05% of rdg +2.0°C)
±(0.1% of rdg +2.5°C)
±(0.1% of rdg +1.5°C)
±3.7°C
±2.7°C
±(0.05% of rdg +2.0°C)
±(0.1% of rdg +2.0°C)
±(0.1% of rdg +2.5°C)
±1.0°C
–100<TS≤800
600<TS≤1820
–100<TS≤100
100<TS≤1100
–100<TS≤1370
B
K
100<TS≤300
R: 300<TS≤1600
S: 300<TS≤1760
400≤TS≤600
Measurement
accuracy
Measurement
accuracy *1
(23°C±5°C)
When 30 minutes
or more have elapsed
after power was
switched on
Filter Line
GND connected
A/D converter
Temperature coefficient
Maximum input voltage
Withstand voltage
Frequency response
Filter
0≤TS≤1300
0≤TS≤2315
E
T
J
N
Type
Number of input channels
Input method
Maximum sampling speed
R/S
0≤TS≤100
Measurement Temperature Range
Measurement Accuracy
Reference contact compensation accuracy
*1: Thermocouple diameters T: 0.32 Φ, others: 0.65 Φ
W
–200≤TS≤–100
–200≤TS≤–100
–200≤TS≤–100
–100<TS≤400
–200<TS≤–100
Input terminal
type
Temperature
Humidity
Voltage
Voltage
Temperature
BNC connectors
All channels isolated
Imbalanced input
Simultaneous sampling of all channels
10Ms
8 channels (fixed)
lThermocouple
20m 50m 100m 200m 500m
1 2 5 10 20 50 100 200 500V 1-5V F.S.
Thermocouple: K, J, E, T, R, S, B, N, W (WRe5-26)
0 to 100% (voltage 0 V to 1 V scaling conversion) *with B-530 (option)
lVoltage
16Bit (out of which 14 bits are internally acknowledged)
Between each input channel terminal/input channel terminal : 60Vp-p
Between each input channel terminal/GND terminal : 60Vp-p
DC to 20KHz (+1/–4dB)
±0.25% of F.S.
M3 screw type terminal board
DescriptionItem
Input Unit Specifications
Gain : 0.01% of F.S./°C
Zero : 0.02% of F.S./°C
Between each input channels and +/- terminals : 20mV to 1V ---> 30Vp-p
: 2V to 500V ---> 500Vp-p
Between each input channel terminal/GND terminal : 1 minute at 1000Vp-p
Between each input channel terminal/input channel terminal : 1 minute at 1000Vp-p
Off, Line, 5Hz, 50Hz, 500Hz
(Attenuation) –3dB / 6dB oct