EasyManua.ls Logo

Greenlee CMP-200 - Page 18

Greenlee CMP-200
68 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
18
Measurements on Three-Phase Circuits (cont’d)
Three-Phase Four-Wire Circuits: Settings Table
*A positive (+) PF indicates an inductive load. A negative (–) PF indicates a capacitive load.
Power factor value is correct for balanced loads only.
**Jaw should be closed and removed from circuit whenever DCA/W ZERO is pressed.
Notes:
The + sign on the jaw must face the power source.
All three power measurements (W
R
, W
S
and W
T
) should be positive. Check test lead
connection and clamping of jaw if any power measurement is negative.
Measurement: True Power, Power Factor*, Apparent Power and Reactive Power
(W, PF, KVA and KVAR)
Type of Circuit: 3ø4W
Selector Setting: 3ø4W
Instructions
Red Lead Black Lead Clamp
Connection Connection Location
R will flash on the display. Press DCA/W ZERO**.
Connect the leads and position the clamp as shown R N R
to take the first measurement (W
R
/ PF
R
).
After the measurement stabilizes, press READ NEXT.
S will flash on the display. Remove clamp from wire.
Press DCA/W ZERO** and connect the leads and S N S
position the clamp as shown to take the final
measurement (W
S
/ PF
S
).
After the measurement stabilizes, press READ NEXT.
T will flash on the display. Remove clamp from wire.
Press DCA/W ZERO** and connect the leads and T N T
position the clamp as shown to take the final
measurement (W
T
/ PF
T
).
After the final measurement stabilizes, press READ NEXT. The CMP-200 will calculate and
display the power factor (PF) and true power (W). Press 3ø KVAR KVA to display the
reactive power (KVAR) and apparent power (KVA).

Related product manuals